首页 | 本学科首页   官方微博 | 高级检索  
     


Compressible flow hot-wire calibration
Authors:R. Johnston  S. Fleeter
Affiliation:(1) School of Mechanical Engineering, Purdue University, West Lafayette, IN 47907, USA, US
Abstract: A resistance thermal anemometer operated in compressible flow is sensitive to density, velocity, and total temperature fluctuations. To address these effects, a relatively inexpensive compressible flow calibration facility has been developed and tested. Characteristics of a hot-film anemometer in compressible flow are determined and a correction technique to account for differences in mean density between the calibration facility and the flow under study is proposed. Received: 19 October 1995 / Accepted: 18 October 1996