Measurement of Er:YAG laser ablation plume dynamics
Authors:J T Walsh Jr  T F Deutsch
Institution:(1) Wellman Laboratories of Photomedicine, Department of Dermatology, Massachusetts General Hospital, 02114 Boston, MA, USA;(2) Present address: Biomedical Engineering Department, Northwestern University, 2145 Sheridan Road, 60208-3107 Evanston IL
Abstract:The dynamics of tissue ablation using an Er:YAG laser were studied using flash photography and optical pump-probe techniques. Both normal-spiking-mode and Q-switched Er:YAG laser radiation were used to study the ablation of skin and bone. Time-resolved photographs of the ablation plume were obtained using a microscope-mounted camera together with pulsed illumination from an excimer-pumped dye laser. The velocity of the plume front, obtained from the photographs, was approximately 1400 m/s. The same velocity was also measured using an optical pump-probe technique. Both techniques indicate that material removal occurred after the end of the 90-ns-long Q-switched laser pulse and that each micropulse in the normal-spiking-mode pulse train was capable of ablating and rapidly ejecting tissue.This work was supported in part by the SDIO-MFEL Program under contract # N00014-86-K-0117 and by the Arthur O. and Gullan M. Wellman Foundation