首页 | 本学科首页   官方微博 | 高级检索  
     


Ultrasensitive Simultaneous Quantification of Nanomolar Level of Cd and Zn by Cathodic Adsorptive Stripping Voltammetry in Some Real Samples
Authors:Shahryar Abbasi  Abbas Farmany  Seyede Shima Mortazavi
Affiliation:1. Department of Chemistry, Ilam University, Ilam, Iran;2. Islamic Azad University, Hamedan Branch, Young Researchers Club, Hamedan, Iran
Abstract:A novel, simple and sensitive adsorptive stripping voltammetry method was developed for simultaneous determination of Cd and Zn using N‐Nitrozo‐N‐phenylhydroxylamine (Cupferron) as a selective complexing agent. Cadmium and zinc metals gave peaks that were distinctly separated by 450–1200 mV, allowing their determination over a wide range of concentrations. The influence of pH and the nature of supporting electrolytes, concentration of ligand, preconcentration time and applied potential were investigated. The detection limits were 0.058 ng/mL for Zn and 0.092 ng/mL for Cd, and the RSD at a concentration level of 50 ppb, were 1.8–2.1 % for both zinc and cadmium, respectively. The method was applied to the determination of cadmium and zinc in blood, drug, food and water samples with the satisfactory results.
Keywords:Simultaneous determination  Adsorptive stripping voltammetry  N‐Nitrozo‐N‐phenylhydroxylamine  Real sample analysis  Nanomolar level
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号