首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Rapid Automatic Determination of Calcium and Magnesium in Aqueous Solutions by FIA Using Potentiometric Detection
Authors:Katja Vahl  Heike Kahlert  Fritz Scholz
Institution:Dept. of Analytical & Environmental Chemistry, Institute of Biochemistry, Greifswald University, Felix‐Hausdorff‐Str. 4, D‐17487 Greifswald, Germany phone: +49‐3834‐864452, fax: +49‐3834‐864451
Abstract:A pH‐sensitive detector for flow‐through potentiometry based on a graphite/quinhydrone/silicone composite electrode was applied to determine calcium and magnesium in aqueous solutions. The determinations are performed by FIA titrations with EDTA as the carrier solution. The method was applied to the analysis of mineral water, drinking water, and river water. The results were compared with those obtained by batch complexometric titration. The proposed new method is suitable for the simple, rapid and automated determination of water hardness in small sample volumes of aqueous solutions. The detection limit for both ions is about 1×10?5 mol L?1, the standard deviation is less than 2%. The sampling frequency is 60 determinations per hour.
Keywords:Flow injection analysis  Flow injection potentiometry  Automatization  Rapid water analysis  Calcium  Magnesium
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号