Effect of viscosity on stationary phase stability in capillary column gas chromatography
Authors:B W Wright  P A Peaden  M L Lee
Abstract:With recent advances in column technology it is now possible to prepare highly efficient, very inert, and thermally stable capillary columns coated with nonpolar polysiloxane stationary phases. Unfortunately, the same degree of success has not been achieved for some of the more polar polysiloxane phases. A parameter that has been studied little in the past in relation to stationary phase film stability is the viscosity of the stationary phase. In this paper the efficiency and stability of coated columns are correlated to the viscosity of the phase. Due to their structure, the viscosity of the phenyl-containing polysiloxanes change rapidly with temperature and hence, thin-film coatings are not stable at elevated temperatures. By using high viscosity phenyl-containing methylphenylpolysiloxanes which were recently synthesized, efficient and stable columns have been prepared.
Keywords:Capillary gas chromatography  Stationary phase stability  Stationary phase viscosity