首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Evaluation of Analytical Countercurrent Chromatographs: High-Speed Countercurrent Chromatograph-4000 Vs. Analytical Toroidal Coil Centrifuge
Abstract:Abstract

Performance of two countercurrent chromatographic models, high speed countercurrent chromatograph (HSCCC-4000) and analytical toroidal coil centrifuge (TCC), is evaluated in terms of theoretical plate number, resolution factor and separation time to assess their analytical capability. A series of experiments was conducted to investigate the effects of internal diameter and length of the coiled column, and flow rate of the mobile phase on the separation of indole auxins in two-phase solvent systems composed of n-hexane-ethyl acetate-methanol-water at different volume ratios. The three components of indole auxins were completely resolved in 16 min with a HSCCC system equipped with a multilayer coil of a 0.55 mm I.D. PTFE tube with theoretical plates ranging from 1290 to 829. Similar separation was achieved in 24 min with a TCC systems equipped with a 0.3 mm I.D. PTFE tube with theoretical plates ranging from 1811 to 969. It is concluded that both systems have comparable analytical capability at the present stage of development.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号