首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A Dual-Column HPLC Method for the Simultaneous Measurement of 6-Thioguanine and Adenine in RNA or DNA
Abstract:Abstract

A sensitive method for measuring 6-thioguanine incorporation into DNA and RNA utilizing a dual column system is presented. The measurement of the 6-thioguanine deoxyribo-or ribonucleosides and deoxyadenosine or adenosine is made simultaneously, thereby allowing for direct calculation of the incorporation per nucleic acid base. The separation utilizes a strong anion-exchange column connected in series with an octadecylsilane column. Prior to high pressure liquid chromatography, the sample is partially purified and oxidized with potassium permanganate. Following a 10-min delay, a 10-min linear gradient from 2% to 20% methanol in 30 mM NH4H2PO4, pH 3.7, is employed. Detection of eluting material is by fluorescence and by UV absorbance at 254 nm. Recovery of the 6-thioguanine nucleosides was determined using 8-14C]-6-thioguanine. The sensitivity of the method for the oxidized 6-thioguanine compounds is approximately 1 pmole (fluorescence) whereas that for the adenine nucleosides (UV absorbance) is about 100 pmoles. This sensitivity is adequate to determine the incorporation in less than 106 (about 1 mg) Chinese hamster ovary cells exposed to a cytotoxic concentration of 6-thioguanine.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号