首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Electrochemical Detection of Retinoids Using Normal Phase HPLC
Abstract:Abstract

Non-aqueous electrochemical (EC) detection of 13-cis-retinoic acid, all-trans retinoic, acitretin and vitamin A palmitate in non-aqueous solvents are reported.

Non-aqueous (EC) detection allows for normal-phase chromato-graphy of these compounds prior to detection. The normal phase system used a mobile phase of HEX/THF/AcOH for the separation of all four compounds. The stationary phase was either silica or PVA-sil. The lipophilic salts, t-butylammoniumtetrafluoroborate or t-butyl-ammoniumhexafluorophosphate necessary for EC detection were added post-column.

The limit of detection (LOD) for EC detection of these compounds is approximately 1 ng on column compared with an LOD by UV absorption of 2 ng on column.

The linear detection for these compounds with the EC detector was about two orders of magnitude.