首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Optimization of background electrolyte composition for simultaneous contactless conductivity and fluorescence detection in capillary electrophoresis of biological samples
Authors:Júlia La n  Jan P&#x;ikryl  Norio Teshima  Hiroya Murakami  Yukihiro Esaka  Franti&#x;ek Foret  Petr Kub&#x;
Institution:Júlia La?ná,Jan P?ikryl,Norio Teshima,Hiroya Murakami,Yukihiro Esaka,Franti?ek Foret,Petr Kubáň
Abstract:In this article, optimization of BGE for simultaneous separation of inorganic ions, organic acids, and glutathione using dual C4D‐LIF detection in capillary electrophoresis is presented. The optimized BGE consisted of 30 mM 2‐4‐(2‐hydroxyethyl)piperazin‐1‐yl]ethanesulfonic acid, 15 mM 2‐amino‐2‐hydroxymethyl‐propane‐1,3‐diol, and 2 mM 18‐crown‐6 at pH 7.2 and allowed simultaneous separation of ten inorganic anions and cations, three organic acids and glutathione in 20 min. The samples were injected hydrodynamically from both capillary ends using the double‐opposite end injection principle. Sensitive detection of anions, cations, and organic acids with micromolar LODs using C4D and simultaneously glutathione with nanomolar LODs using LIF was achieved in a single run. The developed BGE may be useful in analyses of biological samples containing analytes with differing concentrations of several orders of magnitude that is not possible with single detection mode.
Keywords:Capillary electrophoresis  Contactless conductivity detection  Dual detection  Laser induced fluorescence detection  Saliva
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号