首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Selective enrichment and quantification of okadaic acid in shellfish using an immunomagnetic‐bead‐based liquid chromatography with tandem mass spectrometry assay
Authors:Jiaqi Chen  Zhijun Tan  Haiyan Wu  Jixing Peng  Yuxiu Zhai  Mengmeng Guo
Abstract:Okadaic acid is a marine biotoxin that primarily occurs in shellfish and can cause diarrheic shellfish poisoning in humans. When analyzing biological samples using liquid chromatography with tandem mass spectrometry, the presence of complex matrices is a major issue. Thus, it is crucial to selectively and simply extract the target analyte from samples and minimize matrix effects simultaneously. To meet this need, an immunomagnetic‐bead‐based liquid chromatography with tandem mass spectrometry method was developed to detect okadaic acid in shellfish. Magnetic beads bound to monoclonal antibody against okadaic acid were used as affinity probes to specifically enrich okadaic acid in samples, which effectively eliminated matrix effects. A magnetic separator was used to aggregate and separate magnetic particles from sample matrices, and methanol was used to elute okadaic acid from the magnetic beads. Standard solution prepared with methanol was employed directly for quantitative analysis. Several experimental conditions were optimized to improve performance. The method is of interest as a rapid (10 min) sample clean‐up and selective enrichment tool, and it showed good linearity and sensitivity, with reported limits of detection and quantitation of 3 and 10 μg/kg, respectively. Fifty‐three shellfish samples from an aquatic products market were tested using this method, and four samples positive for okadaic acid were found.
Keywords:immunoaffinity  liquid chromatography  okadaic acid  protein G magnetic beads  tandem mass spectrometry
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号