首页 | 本学科首页   官方微博 | 高级检索  
     


Flow Injection Chemiluminescence for the Determination of Estriol via a Noncompetitive Enzyme Immunoassay
Authors:Shuhao Wang  Lingyun Du  Shilei Lin  Huisheng Zhuang
Affiliation:(1) College of Environmental Science and Engineering, Donghua University, 200051 Shanghai, People’s Republic of China;(2) Institute of Chemistry and Chemical Technology, Liaocheng University, 252059 Liaocheng, People’s Republic of China
Abstract:A novel approach to the detection of estriol using a flow injection system coupled to enhanced chemiluminescent immunoassay was developed based on noncompetitive immunoassay formats. A conjugated estriol-ovalbumin immobilized immunoaffinity column was inserted into the flow system to trap the unbound horseradish peroxidase (HRP)-labeled antibody after an off-line incubation of estriol and HRP-labeled anti-estriol antibody. The trapped enzyme conjugate was detected by the injection of chemiluminescent substrates to produce enhanced chemiluminescence. The linear range for the determination of estriol is 10.0 to 400 ng · mL−1 with a correlation coefficient of 0.996 and a detection limit of 5.0 ng · mL−1. The total time for sampling and chemiluminescent detection of one sample is 400 seconds after 30 min of pre-incubation. The results for pregnancy serum samples obtained by this method are in good agreement with those obtained using ELISA.
Keywords:: Flow injection analysis   chemiluminescence   noncompetitive enzyme immunoassay   immunoaffinity   estriol.
本文献已被 SpringerLink 等数据库收录!