首页 | 本学科首页   官方微博 | 高级检索  
     


The effect of electric field strength,buffer type and concentration on separation parameters in capillary zone electrophoresis
Authors:H. J. Issaq  I. Z. Atamna  G. M. Muschik  G. M. Janini
Affiliation:(1) PRI/Dyn Corp, NCI-Frederick Cancer Research and Development Center, P.O. Box B, 21702-1201 Frederick, Maryland, USA
Abstract:Summary Probe solutes were used to investigate the effect of buffer type, concentration and applied voltage on solute mobility, column efficiency and resolution in capillary zone electrophoresis. With low conductivity buffers higher concentrations and/or higher voltages could be used to improve column efficiency and resolution. Doubling the concentration of the buffer doubles the amount of heat generated inside the column while doubling the applied voltage cause a 4-fold increase. Solute migration time is approximately an inverse function of the charge density of the buffer's cation. Analysis time is increased by about 30% if the buffer concentration is doubled while it is cut in half if the applied voltage is doubled. Column efficiency is improved (higher theoretical plate count) with increasing buffer concentration and/or applied voltage as long as the heat generated is efficiently dissipated. The separation factor is directly related to analysis time and, therefore, selectivity improves with increasing buffer concentration but decreases with increasing applied voltage. Hence, resolution is optimized by increasing buffer concentration at a moderate applied voltage.
Keywords:Capillary zone electrophoresis  Optimization of resolution  Applied voltage  Buffer concentration  Buffer type  Electroosmotic mobility  Electrophoretic mobility
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号