Analysis of adenosine by RP-HPLC method and its application to the study of adenosine kinase kinetics
Authors:Marin Rodrigo Miguel  Franchini Kleber Gomes  Rocco Silvana Aparecida
Institution:Department of Internal Medicine, School of Medicine, UNICAMP, Campinas, SP, Brazil.
Abstract:An RP-HPLC method for the analysis of adenosine (ADO) has been developed and validated. In the present study, we report an RP-HPLC-based method with modifications of mobile phase and shorter retention time that substantially improved the efficiency of ADO analysis. The HPLC separation of the ADO was achieved on a C18 column, using a mobile phase consisting of water, containing 7% v/v ACN, at a flow rate of 0.8 mL/min. The column effluent was monitored by UV detection at 260 nm. A linear response was achieved over the concentration range of 0.25-100.00 micromol/L. The analytical method inter- and intra-run accuracy and precision were better than +/- 15%. The LOQ was 0.25 micromol/L, with ADO detection in the range of 6.25 pmol per sample. The method has been applied to the study of adenosine kinase (AK) kinetics.