Simultaneous measurements of multiple flow velocity components using frequency modulated lasers and a single molecular absorption cell
Authors:Andreas Fischer,Lars Bü  ttner,Jü  rgen Czarske
Affiliation:
  • Laboratory for Measurement and Testing Techniques, Technische Universität Dresden, Helmholtzstr. 18, 01062 Dresden, Germany
  • Abstract:Since unsteady, complex flow phenomena play an important role, optical measurements techniques are required for flow investigations, which provide simultaneous measurements of multiple velocity components. Doppler global velocimetry has this potential. It is a flow measurement technique, where the Doppler shift of scattered light is measured by a molecular absorption cell for frequency-to-intensity conversion. However, novel Doppler global velocimeters with laser frequency modulation were only used for single component measurements yet. In order to enhance such a system for the simultaneous measurement of multiple components, a concept based on frequency division multiplexing is introduced for the first time. Besides multiple lasers, only a single molecular absorption cell and a single detector unit is required. Two-component measurements of velocity profiles from nozzle flows are presented as a proof of principle. The designed measurement system provides high measurement rates of up to 20 kHz, which is three orders of magnitude higher than for typical Doppler global velocimetry setups.
    Keywords:Diode laser   Frequency modulation   Flow velocity   Doppler effect
    本文献已被 ScienceDirect 等数据库收录!
    设为首页 | 免责声明 | 关于勤云 | 加入收藏