首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High dynamic range imaging method for interferometry
Authors:J Vargas  R Restrepo  T Belenguer
Institution:
  • a Laboratorio de Instrumentación Espacial, Instituto Nacional de Técnica Aeroespacial, Carretera de Ajalvir Km 4, 28850, Torrejón de Ardoz (Madrid), Spain
  • b Optics Department, Universidad Complutense de Madrid, Facultad de CC. Físicas, Ciudad Universitaria s/n, 28040 Madrid, Spain
  • Abstract:We demonstrate a method to easily and quickly extend the dynamic range imaging capabilities of the camera in a typical interferometric approach. The camera dynamic range is usually low and limited to 256 gray levels. Also, it is well known that one may have over or under-exposed regions in the interferogram (due to non-uniform illumination) which makes these image regions not reliable. In our proposed method it is not necessary to obtain or use the non-linear camera response curve in order to extend the camera dynamic range. We obtain a sequence of differently exposed interferograms, typically five or six; after that, we compute the corresponding normalized fringe patterns and modulation maps using a typical normalization method. These normalized patterns are combined through a temporal weighted average using as weights the corresponding modulation maps. We show a set of experimental results that prove the effectiveness of the proposed method.
    Keywords:High dynamic range imaging  Interferometry  Fringe analysis