首页 | 本学科首页   官方微博 | 高级检索  
     检索      


High-throughput biophysical measurement of human red blood cells
Authors:Zheng Yi  Shojaei-Baghini Ehsan  Azad Azar  Wang Chen  Sun Yu
Institution:Department of Mechanical and Industrial Engineering, University of Toronto, Toronto, ON M5S 3G8, Canada.
Abstract:This paper reports a microfluidic system for biophysical characterization of red blood cells (RBCs) at a speed of 100-150 cells s(-1). Electrical impedance measurement is made when single RBCs flow through a constriction channel that is marginally smaller than RBCs' diameters. The multiple parameters quantified as mechanical and electrical signatures of each RBC include transit time, impedance amplitude ratio, and impedance phase increase. Histograms, compiled from 84,073 adult RBCs (from 5 adult blood samples) and 82,253 neonatal RBCs (from 5 newborn blood samples), reveal different biophysical properties across samples and between the adult and neonatal RBC populations. In comparison with previously reported microfluidic devices for single RBC biophysical measurement, this system has a higher throughput, higher signal to noise ratio, and the capability of performing multi-parameter measurements.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏