Design study of hard X-ray tomography system to obtain a spatial resolution of 100 nm
Authors:Kwon Su ChonSeon Kwan Juhng  Kwon-Ha Yoon
Affiliation:a Department of Radiological Science, Catholic University of Daegu, 333 Kumrakro, Hayang, Gyeongsan, Gyeongbuk 712-702, Republic of Korea
b Institute for Radiological Imaging Science, Wonkwang University, 344-2 Sinyong-dong, Iksan, Jeonbuk 570-749, Republic of Korea
c Radiation Imaging Technology Center, Jeonbuk Technopark, 500 Dasongri, Iksan, Jeonbuk 570-802, Republic of Korea
Abstract:An optic-based X-ray tomography system of a high spatial resolution using a conventional X-ray tube was proposed. The system had several X-ray optics: multilayer mirror for monochromatic X-ray, capillary optic for focusing X-ray onto a sample, and objective zone plate. The X-ray tomography system was designed for obtaining a spatial resolution of 100 nm. Design parameters for each optic were determined and optimized by ray tracing in considering X-ray intensity and reflectivity. The X-ray tomography system with a spatial resolution of 100 nm will provide a good inspect tool in bio-medical field and semiconductor applications.
Keywords:X-ray tomography system   X-ray optics   Spatial resolution
本文献已被 ScienceDirect 等数据库收录!