首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Headspace gas chromatography analysis of toxic contaminants in ethoxylated alcohols and alkylamines
Authors:A Wala-Jerzykiewicz  J Szymanowski
Institution:(1) Institute of Heavy Organic Chemicals, ul. Energetyków 9, 47-225 Kędzierzyn-Koźle, Poland;(2) Institute of Chemical Technology and Engineering, Poznań University of Technology, Pl. Skłodowskiej-Curie 2, 60-965 Poznań, Poland
Abstract:Summary Headspace-gas chromatography was used to determine the contents of toxic 1,4-dioxane, ethylene oxide and ethylene glycol in ethoxylated alcohols and alkylamines, and in commercial cosmetics and washing products. A Permaphase PEG capillary column was used for the determination of 1,4-dioxane and ethylene oxide and a DB-17 column for ethylene glycol determination. Dimethylformamide was used as the solvent in the determination of 1,4-dioxane and ethylene oxide, and undecanol in the case of ethylene glycol. The detection limits for ethylene oxide, 1,4-dioxane and ethylene glycol are 1,2 and 10 μg·g−1, respectively.
Keywords:Gas chromatography  Headspace sampling  Ethoxylated alcohols  Ethoxylated alkylamines
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏