Improvement of vacuum nebulizing interface for direct coupling micro-liquid chromatograph with mass spectrometer and some applications to polar natural organic compounds |
| |
Authors: | H. Yoshida K. Matsumoto K. Itoh S. Tsuge Y. Hirata K. Mochizuki N. Kokubun Y. Yoshida |
| |
Affiliation: | (1) Department of Synthetic Chemistry, Faculty of Engineering, Nagoya University, 464 Nagoya, Japan;(2) School of Materials Science, Toyohashi University of Technology, 440 Toyohashi, Japan;(3) JASCO, Ishikawa-cho, 2967-5 Hachioji, 192 Tokyo, Japan |
| |
Abstract: | Summary A vacuum nebulizing interface for direct coupling a micro-liquid chromatograph (LC) with a mass spectrometer (MS) was improved by incorporating a small water cooling jacket and a bubble saturator for the nebulizing gas. By this improvement the applicability of the micro-LC-MS system was remarkably extended to various non-volatile polar organic compounds such as adenosine, tryptophane, glycyl-glycyl-glycine, mono- and di-saccharides, etc. The desirable temperature profile around the nebulizing tip was also discussed.
Verbesserung eines Vakuumzerstäubers als Zwischenstück zur direkten Kopplung von Mikro-Flüssig-Chromatograph und Massenspektrometer. Anwendung auf polare organische Verbindungen Zusammenfassung Ein Vakuumzerstäuber wurde durch Einführung eines Wasser-Kühlmantels und einer Sättigungsvorrichtung für das Zerstäubergas verbessert. Dadurch konnte die Anwendbarkeit des Systems auf verschiedenartige nichtflüchtige polare organische Verbindungen (u. a. Adenosin, Tryptophan, Glycyl-glycyl-glycin, Mono- und Disaccharide) ausgedehnt werden. Das optimale Temperaturprofil und die Zerstäuberspitze wurde ebenfalls untersucht. |
| |
Keywords: | |