Extrem nachweisstarkes gas-chromatographisches Bestimmungsverfahren für Selen nach Abtrennung durch Verdampfen im Sauerstoffstrom |
| |
Authors: | A. Meyer E. Grallath G. Kaiser G. Tölg |
| |
Affiliation: | 1. Laboratorium für Reinststoffe, Stuttgart und Schw?bisch Gmünd, Max-Planck-Institut für Metallforschung, Deutschland
|
| |
Abstract: | For the determination of the smallest amounts of selenium in high-purity copper, the sample (≤2 g) is heated in a quartz tube at 1100–1150°C. SeO2 evaporates and is collected quantitatively by condensation of the carrier gas in a micro-trap cooled with liquid nitrogen. After evaporation of the oxygen, SeO2 is caused to react in the trap with 4-nitro-o-phenylenediamine, forming 5-nitropiazselenol, which is then extracted with toluene and determined in the ppb-range by gas-chromatography with ECD. All separation steps were optimized by the tracer method using 75Se. Determining 0.74 ppm of Se, the standard deviation was found to be 0.004 ppm for 15 determinations with samples of 200–600 mg. The detection limit is 1 ppb of Se. |
| |