首页 | 本学科首页   官方微博 | 高级检索  
     


One-step electropolymeric co-immobilization of glucose oxidase and heparin for amperometric biosensing of glucose
Authors:Wang J  Chen L  Hocevar S B  Ogorevc B
Affiliation:Department of Chemistry and Biochemistry, New Mexico State University, Las Cruces 88003, USA.
Abstract:Electropolymerization was used for the co-deposition of glucose oxidase and heparin onto metal electrode transducers. Such electropolymeric co-entrapment within a non-conducting poly(1,2-phenylenediamine) (PPD) film imparts both biocatalytic and anticoagulation activities onto the transducer, and greatly improves the performance of the sensor after exposure to whole blood. Essentially identical glucose signals are observed before and after exposure to blood samples. Scanning electron micrographs after such exposure reveal no platelet deposition or formation of a fibrin "clot". The effect of the heparin co-immobilization on the glucose response is examined. Improved biocompatibility is reported also in connection with a needle-type carbon paste biosensor configuration. The simultaneous localization of the enzyme and heparin offers great promise for simplifying the preparation of enzyme electrodes and designing biocompatible implantable glucose biosensors.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号