首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Carbon nanotube wiring for signal amplification of electrochemical magneto immunosensors: application to myeloperoxidase detection
Authors:Zorione Herrasti  Fernando Martínez  Eva Baldrich
Institution:1. IK4-Ikerlan Technological Research Centre, 20500, Mondragón, Spain
2. Institut de Microelectrònica de Barcelona (IMB-CNM, CSIC), Esfera UAB, Campus Universitat Autònoma de Barcelona, 08193, Bellaterra, Spain
3. Diagnostic Nanotools, Molecular Biology and Biochemistry Research Center for Nanomedicine (Cibbim-Nanomedicine), Vall d’Hebron Institut de Recerca (VHIR), Universitat Autònoma de Barcelona, Passeig de la Vall d’Hebron 119-129, 08035, Barcelona, Spain
Abstract:In this work, chronoamperometric myelo-peroxidase (MPO) detection was accomplished using immunofunctionalized magnetic microparticles (MPs), disposable carbon screen-printed electrodes (C-SPEs), and a ready-to-use commercially available tetramethylbenzidine (TMB)-based enzymatic substrate. In order to reach the limit of detection (LOD) needed to study real blood serum samples, assay performance was additionally improved by exploiting CNT wiring, which amplified the signal and decreased the LOD. The optimized assay can be performed in 30 min and yields LODs of 6 and 55 ng mL?1 in PBS and undiluted human serum, respectively, making it useful for the identification of patients at risk of cardiovascular disease. These results demonstrate that electrode nanostructuring can be accomplished “post-assay,” which favors the development of enhanced magneto immunosensors based on the exploitation of cheap and simple SPE devices.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号