Abstract: | An optical Doppler tomography (ODT) system that permits imaging of fluid flow velocity in highly scattering media is described. ODT combines Doppler velocimetry with the high spatial resolution of low-coherence optical interferometry to measure fluid flow velocity at discrete spatial locations. Tomographic imaging of particle flow velocity within a circular conduit submerged 1 mm below the surface in a highly scattering phantom of Intralipid is demonstrated. |