首页 | 本学科首页   官方微博 | 高级检索  
     


Pulsed Electron Spin Resonance Ex Situ Probe for Tooth Biodosimetry
Authors:Ksenia Sirota  Ygal Twig  Aharon Blank
Affiliation:1. Schulich Faculty of Chemistry, Technion, Israel Institute of Technology, 32000, Haifa, Israel
Abstract:An electron spin resonance (ESR) probe that includes a static field source and a microwave resonator for the measurement of paramagnetic defects in tooth enamel is presented. Such defects are known to be a good marker for quantifying the amount of ionizing radiation dose absorbed in the tooth. The probe can measure the tooth when it is positioned just above its outer surface, i.e., in ex situ geometry. It is operated in pulsed mode at a frequency of ~6.2 GHz that corresponds to the magnitude of the static magnetic field of its permanent magnet. A detailed design of the probe is provided, together with its specifications in terms of measurement volume and signal-to-noise-ratio for a typical sample. Experimental results that verify its sensitivity and capability to measure gamma-irradiated teeth are provided. The current minimal detected signal by the probe corresponds to a radiation dose of ~4 Gy.
Keywords:
本文献已被 SpringerLink 等数据库收录!