首页 | 本学科首页   官方微博 | 高级检索  
     

密闭舱内VOCs现场快速定量检测技术与装置研究
引用本文:李盛红,丁坤,孟虎,关亚风. 密闭舱内VOCs现场快速定量检测技术与装置研究[J]. 分析测试学报, 2021, 40(4): 596-604
作者姓名:李盛红  丁坤  孟虎  关亚风
作者单位:中国科学院大连化学物理研究所,辽宁大连116023
基金项目:大连市青年科技之星项目(2019RQ056);中国科学院战略性先导科技专项(A类)子子课题(XDA22040405-02)
摘    要:该文发展了可对压强变化的密封舱实现计量采样-富集的采样技术,基于薄壳金属筒式低功耗均温色谱柱组件,结合微池热导检测器/小型氢火焰离子化检测器,建立了一种对密闭舱内挥发性有机物(VOCs)进行现场快速定量检测的方法和装置.根据负压罐和密封舱的压强差值和绝压值以及温度,计算出有效采样体积(折算到标准大气压),进而计算出富集...

关 键 词:密闭舱  VOCs  现场快速检测  计量采样技术  低功耗色谱柱

Study on an On-site Rapid Quantitative Analysis Device for Volatile Organic Compounds in Sealed Cabin
LI Sheng-hong,DING Kun,MENG Hu,GUAN Ya-feng. Study on an On-site Rapid Quantitative Analysis Device for Volatile Organic Compounds in Sealed Cabin[J]. Journal of Instrumental Analysis, 2021, 40(4): 596-604
Authors:LI Sheng-hong  DING Kun  MENG Hu  GUAN Ya-feng
Affiliation:Dalian Institute of Chemical Physics,Chinese Academy of Sciences,Dalian116023,China
Abstract:In this paper,an on-site rapid quantitative analysis device of volatile organic compounds(VOCs)in sealed cabin was developed.The device was based on the metrological sampling-enrichment technique with thin-shell metal cylinder type capillary column heater,and micro thermal conductivity detector/micro flame ionization detector.To adapt pressure change in sealed cabin,a fix-volume vacuum tank with a pressure sensor and a temperature sensor was used to calculate the effective sampling volume and the enrichment factor.The enrichment factor was 400 for toluene at an effective sampling volume of 100 mL(converted to standard atmospheric pressure).The column assembly was constructed by wrapping the heating wire and column on the inside and outside layers of a thin-shell metal cylinder,respectively.The temperature uniformity of capillary column was less than 0.4℃,and a temperature programming rate of 40℃/min was obtained.In the heating process from 40℃to 300℃at 10℃/min,the power consumption was less than 35 W,while at an isothermal temperature of 300℃,the power consumption was only 28 W.The separation performances,including peak width at half height,column efficiency,resolution and repeatability of the column assembly were evaluated and compared with a laboratory chromatograph,proving that the separation performances of the two heating methods were similar.The sampling-enrichment performance was evaluated with varieties of VOCs samples.In the high speed separation mode,53 VOCs could be separated within 5 minutes using a peak width at half height less than 0.8's for all analytes.The developed device was applied to the rapid determination of hydrocarbons,benzene homogeneous,alcohols,aldehydes,ketones and other VOCs.The performance parameters of the device were satisfactory,compared with those of commercial portable gas chromatography(GC)or gas chromatography-mass spectrometry(GC-MS).
Keywords:sealed cabin  volatile organic compounds(VOCs)  on site rapid analysis,metrological sampling technique  low power consumption column assembly
本文献已被 维普 等数据库收录!
点击此处可从《分析测试学报》浏览原始摘要信息