首页 | 本学科首页   官方微博 | 高级检索  
     


Accuracy in the determination of the compositions of main rockforming silicates and oxides on a JXA-8100 microanalyzer
Authors:V. N. Korolyuk  L. V. Usova  E. N. Nigmatulina
Affiliation:1.Institute of Geology and Mineralogy, Siberian Branch,Russian Academy of Sciences,Novosibirsk,Russia
Abstract:The individual sources of errors in procedures used for considering matrix effects were studied. The ZAF and PRZ analytical programs provided good correction in the predominance of one of the matrix effects, namely, absorption or the atomic number. Their simultaneous occurrence increased the total error to 2–4 rel %. In the ZAF method, normalization to oxygen in terms of stoichiometry introduced errors if the sample simultaneously contained the test element in various oxide forms. The greatest error of up to 1–2 wt % appeared in simple and double oxides. In the analysis of the main rockforming silicates with the determination of total iron as FeO, the presence of Fe2O3 can change the concentration of components to ∼0.2%.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏