首页 | 本学科首页   官方微博 | 高级检索  
     


A pulsed-leak valve for use with ion trapping mass spectrometers
Authors:Charles Q. Jiao  Don Rufus A. Ranatunga  Weldon E. Vaughn  Ben S. Freiser
Affiliation:1. H.C. Brown Laboratory of Chemistry, Purdue University, West Lafayette, Indiana, USA
Abstract:
A pulsed-leak valve that allows the introduction of a prolonged, flat, and controllable pulse of gas is described. Test results from the valve that utilized a Fourier transform ion cyclotron resonance mass spectrometer with Ar and C2H6 as the sample gases indicate that the valve functions as expected and yields basically rectangular pressure profiles in the cell region. The rise and fall times are similar to those of just the stand-alone pulsed valve and are believed to be determined mainly by the design of the vacuum system, rather than the design of the pulsed-leak valve. Kinetic data for the reaction of Nb+ with C2H6, acquired with the use of the pulsed-leak valve to introduce the C2H6 gas, demonstrates the practical application of this valve for kinetic and other analogous studies. Use of the pulsed-leak valve greatly reduces the loss of the reactant ion signal during the cooling period.
Keywords:
本文献已被 ScienceDirect SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号