Optimization of the atmospheric pressure chemical ionization liquid chromatography mass spectrometry interface
Authors:D M Garcia  S K Huang  W F Stansbury
Institution:1. Analytical Department, Hoechst Celanese Corporation, 1901 Clarkwood Road, 78469, Corpus Christi, TX
Abstract:Use of optimized instrument parameters that result from statistical experimentation revealed that the sensitivity of atmospheric pressure chemical ionization (APCI) liquid chromatography-mass spectrometry (LC/MS) is greater than the sensitivity of an optimized Thermabeam? LC/MS interface by about 3 orders of magnitude, when tested on aromatic compounds. APCI is one of the few LC/MS techniques in which the chromatogram is directly comparable with liquid chromatographs that use ultraviolet detection. The optimum instrument parameters for a Finnigan SSQ-7000 APCI LC/MS interface were found at low flow rates (e. g., 0. 1 mL/min), relatively low capillary heat (e. g., 225 °C), and high sheath-gas pressure (e. g., 60 lb/in2). The optimization was achieved by monitoring the responses of sensitivity, fragmentation, and cluster ion formation. The fine tuning for high sensitivity calls for a high percentage of water in the mobile phase. In contrast, a high percentage of organic content in the mobile phase is required to obtain abundant protonated molecular ions with respect to fragmentation and clustering. This is an important consideration for analyses of unknowns.