Head-column field-amplified sample stacking in a capillary electrophoresis-flow injection system |
| |
Authors: | Fan Liuyin Cheng Yuqiao Li Yuqin Chen Hongli Chen Xingguo Hu Zhide |
| |
Institution: | Department of Chemistry, Lanzhou University, PR China. |
| |
Abstract: | A simple, effective, and continuous online concentration method for the sensitive detection of alkaloids applying CE-flow injection analysis with head-column field-amplified sample stacking was developed. A series of samples was continuously introduced into the capillary by electrokinetic means without interrupting the high voltage. A short water plug was introduced by the EOF at the capillary inlet end prior to sample introduction. Under optimum conditions, 15-fold improvement in concentration sensitivity was achieved, giving an LOD of about 0.67 and 0.73 microg/mL for ephedrine (E) and pseudoephedrine (PE), respectively. The separation could be achieved within 4 min and sample throughput rate could reach up to 7/h. The repeatability (defined as RSD) was 3.62, 1.51% with peak area evaluation and 1.30, 2.58% with peak height evaluation for E and PE, respectively. This method has been successfully applied to the analysis of commercial pharmaceutical preparations containing E and PE, and the recoveries were 92.3-102.4%. |
| |
Keywords: | Capillary electrophoresis Flow injection Head‐column field‐amplified sample stacking |
本文献已被 PubMed 等数据库收录! |
|