首页 | 本学科首页   官方微博 | 高级检索  
     


Quantification of pramipexole in human plasma by liquid chromatography tandem mass spectrometry using tamsulosin as internal standard
Authors:Nirogi Ramakrishna V S  Kandikere Vishwottam  Shrivastava Wishu  Mudigonda Koteshwara  Maurya Santosh  Ajjala Devender
Affiliation:Biopharmaceutical Research, Suven Life Sciences Ltd, Serene Chambers, Road No. 5, Avenue 7, Banjara Hills, Hyderabad, India. Ramakrishna_nirogi@yahoo.co.in
Abstract:A high-performance liquid chromatography/electrospray ionization tandem mass spectrometry method was developed and validated for the quantification of pramipexole in human plasma. Following liquid-liquid extraction, the analytes were separated using an isocratic mobile phase on a reverse-phase column and analyzed by MS/MS in the multiple reaction monitoring mode using the respective [M + H](+) ions, m/z 212/152 for pramipexole and m/z 409/228 for the IS. The method exhibited a linear dynamic range of 200-8000 pg/mL for pramipexole in human plasma. The lower limit of quantification was 200 pg/mL with a relative standard deviation of less than 8%. Acceptable precision and accuracy were obtained for concentrations over the standard curve range. A run time of 3.5 min for each sample made it possible to analyze more than 200 human plasma samples per day. The validated method has been successfully used to analyze human plasma samples for application in pharmacokinetic, bioavailability or bioequivalence studies.
Keywords:pramipexole  liquid chromatography–tandem mass spectrometry  human plasma  pharmacokinetic study
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号