Multisyringe flow injection analysis system for automation of standard addition calibration method
Authors:Joã  o Rodrigo Santos,Marcela A. Segundo,José   L.F.C. Lima,Mauro Korn
Affiliation:aREQUIMTE, Serviço de Química-Física, Faculdade de Farmácia, Universidade do Porto, R. Aníbal Cunha, 164, 4099-030 Porto, Portugal;bDept. de Ciências Exactas e da Terra, Universidade do Estado da Bahia, R. Silveira Martins, 2555, 41195-001, Salvador, BA, Brasil
Abstract:In the present work, the multi-channel features of multisyringe flow injection analysis (MSFIA) were exploited for the first time to implement calibration based on standard addition method (SAM). For this, standard solutions containing different concentrations of target analyte were placed in each syringe of the multisyringe and connected to a flow network where in-line mixing of sample and standard through a merging zone approach was established prior to detection of analyte. Using this strategy, artifacts reported before in SAM using flow injection analysis were avoided as the concentration of the analyte in the resulting mixture was related to the dilution of sample and added standard within the system, and the concentration of all matrix components was kept constant during all measurements. The feasibility of the proposed MSFIA system was assessed through application to potentiometric determination of chloride ion in electroplating bath and milk samples. Results obtained for samples (n = 15) were not statistically different from those provided by titrimetric procedures, with an excellent throughput (20–31 samples h− 1), comprising four-level addition of chloride ion.
Keywords:Standard addition method   Multisyringe flow injection analysis   Potentiometry   Chloride determination   Electroplating bath   Milk
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号