Abstract: | This paper reports two subjects. It describes LC-GC transfer by partially concurrent eluent evaporation at a greatly accelerated rate, as required for optimal compatibility with 2–3 mm i.d. LC columns and LC flow rates up to some 500 μl/min. Evaporation rates around 200 μl/min were obtained using a 0.53 mm. i.d. uncoated pre-column and an early vapor exit. A stationary-phasecoated “retaining” pre-column was used for preventing escape of volatile solutes through the vapor exit. The technique was used for the detection of food irradiation by analyzing selected radiolysis products of triglycerides, namely alkanes/alkanes and aldehydes. Extracted fat of chicken, hazel-nuts, and soup mixes was injected in LO and the relevant fractions were transferred on-line to GC. For chicken and nuts, detection of irradiation was possible down to doses below 0.5 kGy. Detection limits were higher for soup mixes due to interfering peaks. |