Ruthenium/rhodium modified gold electrodes for the amperometric detection of hydrogen peroxide at low potentials
Authors:Janasek Dirk  Vastarella Walter  Spohn Uwe  Teuscher Nico  Heilmann Andreas
Institution:Institute of Biotechnology, Martin-Luther-University Halle-Wittenberg, Kurt-Mothes-Strasse 3, Germany.
Abstract:Electrodes of ruthenium/rhodium deposited as thin layers on gold foils were investigated. Ruthenium layers were radio frequency (r.f.) magnetron sputtered and the rhodium layers were made by vacuum evaporation. Hydrogen peroxide could be detected using the cathodic reduction at potentials lower than +170 mV or the anodic oxidation at higher potentials. Under flow injection conditions, H(2)O(2) was detected between 1 and 1000 micro M at a potential of -100 mV and between 2 and 500 micro M at a potential of +250 mV vs. Ag/AgCl/0.4 M KCl. The electrodes also showed high operational stability and selectivity against many electroactive substances. The selectivity against dissolved oxygen was investigated.