Collecting efficiency of a cylindrical mirror electron energy analyzer with pre-retarding lens
Authors:J.L. Gardner  James A.R. Samson
Affiliation:Behlen Laboratory of Physics, University of Nebraska, Lincoln, Nebraska 68508 U.S.A.
Abstract:The electron collecting efficiency of a cylindrical mirror energy analyzer incorporating retardation of the electrons prior to analysis has been determined over the range 0 to 30 eV by two methods. The first method requires the use of a vacuum ultraviolet monochromator to produce monoenergetic electrons of different energies; the second method involves measuring the energy-brightness relationship of the retarding optics and should be applicable to any deflection analyzer with pre-retarding optics. The results of the two methods are compared and the limitations of the latter method are discussed.