Field Measurement of NO2 and RNO2 by Two-Channel Thermal Dissociation Cavity Ring Down Spectrometer
Authors:Jian Chen  Hao Wu  An-wen Liu  Shui-ming Hu  Jingsong Zhang
Institution:1.Hefei National Laboratory for Physical Sciences at the Microscale, University of Science and Tech-nology of China, Hefei 230026, China;Department of Chemistry and Air Pollution Research Center, University of California, Riverside, California 92521, America2.Hefei National Laboratory for Physical Sciences at the Microscale, University of Science and Tech-nology of China, Hefei 230026, China
Abstract:A two-channel thermal dissociation cavity ring down spectroscopy (CRDS) instrument has been built for in situ, real-time measurement of NO2 and total RNO2 (peroxy nitrates and alkyl nitrates) in ambient air, with a NO2 detection limit of 0.10 ppbv at 1 s. A 6-day long measurement was conducted at urban site of Hefei by using the CRDS instrument with a time resolution of 3 s. A commercial molybdenum converted chemiluminescence (Mo-CL) instrument was also used for comparison. The average RNO2 concentration in the 6 days was measured to be 1.94 ppbv. The Mo-CL instrument overestimated the NO2 concentration by a bias of +1.69 ppbv in average, for the reason that it cannot distinguish RNO2 from NO2. The relative bias could be over 100% during the afternoon hours when NO2 was low but RNO2 was high.
Keywords:Cavity ring-down spectroscopy  Molybdenum converted chemiluminescence  Thermal dissociation  Organic nitrates