首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Temperature monitoring using ultrasound contrast agents: in vitro investigation on thermal stability
Authors:Guiot C  Cavalli R  Gaglioti P  Danelon D  Musacchio C  Trotta M  Todros T
Institution:Department of Neuroscience, University of Torino, Physiology Section, C. Raffaello 30, 10125 Torino, Italy. caterina.guiot@unito.it
Abstract:Invasiveness of temperature monitoring devices is presently one of the most serious limitations to the application of oncological hyperthermia (HT). A promising approach aims at detecting temperature variations by monitoring the mean grey level (MGL) of the ultrasonographic image of the tissue. Gaseous ultrasound contrast agents (UCA), enhancing Ultrasonic (US) imaging, are expected to be sensitive to temperature, and are therefore a good candidate as temperature monitoring medium. The present study evaluates the 'in vitro' temporal and thermal stability and the correlation between temperature and MGL using a gaseous UCA (SonoVue) as phantom. No statistical differences were detected between the MGL value of the phantom kept at 43.5 degrees C before (215.2+/-3.5) and after 1 h (214.8+/-2.5), showing good stability at HT temperatures. Data of MGL image vs. temperature were obtained during both heating and cooling experiments in the HT range (30-43 degrees C). A good linearity of MGL vs. temperature (R2=0.976) was found with a good accuracy (2.5%) and a sensitivity of about 6.6 MGL/degrees C.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号