Fast and direct determination of butylated hydroxyanisole in biodiesel by batch injection analysis with amperometric detection |
| |
Authors: | Tormin Thiago Faria Gimenes Denise Tofanello Richter Eduardo Mathias Munoz Rodrigo Alejandro Abarza |
| |
Affiliation: | Instituto de Química, Universidade Federal de Uberlândia, Av. João Naves de Ávila, 2121, Uberlândia, MG, Brazil |
| |
Abstract: | We report here, for the first time, application of batch injection analysis (BIA) with amperometric detection for determination of the phenolic antioxidant butylated hydroxyanisole (BHA) in biodiesel. A sample plug was directly injected onto a boron-doped diamond electrode immersed in 50% v/v hydroethanolic solution with 0.1 mol L−1 HClO4 using an electronic micropipette. Importantly, the only preparation step required for biodiesel analysis is dilution in the same hydroethanolic electrolyte solution. Our proposed method has several advantages for routine biodiesel analysis, including: a low relative standard deviation between injections (0.29%, n = 20), high analytical frequency (120 h−1), adequate recovery values (93-101%) for spiked samples, satisfactory accuracy (based on comparative determinations by high-performance liquid-chromatography), and a low detection limit (100 ng of BHA per g of biodiesel). Finally, our method can be adapted for the determination of other antioxidants in biodiesel samples. |
| |
Keywords: | Batch injection analysis Amperometry Biodiesel Antioxidants Boron-doped diamond electrode |