Use of multi-coil parallel-gap resonators for co-registration EPR/NMR imaging
Authors:Kawada Yuuki  Hirata Hiroshi  Fujii Hirodata
Institution:Department of Electrical Engineering, Yamagata University, Yonezawa, Yamagata 992-8510, Japan.
Abstract:This article reports experimental investigations on the use of RF resonators for continuous-wave electron paramagnetic resonance (cw-EPR) and proton nuclear magnetic resonance (NMR) imaging. We developed a composite resonator system with multi-coil parallel-gap resonators for co-registration EPR/NMR imaging. The resonance frequencies of each resonator were 21.8MHz for NMR and 670MHz for EPR. A smaller resonator (22mm in diameter) for use in EPR was placed coaxially in a larger resonator (40mm in diameter) for use in NMR. RF magnetic fields in the composite resonator system were visualized by measuring a homogeneous 4-hydroxy-2,2,6,6-tetramethyl-piperidinooxy (4-hydroxy-TEMPO) solution in a test tube. A phantom of five tubes containing distilled water and 4-hydroxy-TEMPO solution was also measured to demonstrate the potential usefulness of this composite resonator system in biomedical science. An image of unpaired electrons was obtained for 4-hydroxy-TEMPO in three tubes, and was successfully mapped on the proton image for five tubes. Technical problems in the implementation of a composite resonator system are discussed with regard to co-registration EPR/NMR imaging for animal experiments.