The use of calix[6]arene molecules for actinides analysis in urine and drinking water: an alternative to current procedures |
| |
Authors: | C. Bouvier-Capely A. Manoury A. Legrand J. P. Bonthonneau F. Cuenot F. Rebière |
| |
Affiliation: | (1) Radiological Protection and Human Health Division, IRSN, B.P. 17, 92 262 Fontenay-aux-Roses Cedex, France |
| |
Abstract: | In this work the applicability of calix[6]arene columns for actinides analysis in urine samples and drinking water was investigated. A radiochemical procedure has been developed for U, Pu, Am analysis in urine. A simple and effective method has also been proposed on a specific column named AQUALIX, for the separation and preconcentration of U from drinking water. These procedures are suitable for routine analysis and require a considerably reduced number of steps of sample treatment as compared to current procedures. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|