Comparison of three electrochemical end-point detection methods to assay potassium dichromate by coulometric titration |
| |
Authors: | Toshiaki Asakai Akiharu Hioki |
| |
Affiliation: | (1) National Metrology Institute of Japan (NMIJ), National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba Central 3, 1-1-1 Umezono, Tsukuba Ibaraki, 305-8563, Japan |
| |
Abstract: | Coulometric titrations with three electrochemical end-point detection methods were performed to assay potassium dichromate as a standard for oxidation–reduction titration. The assay as an oxidizing agent was carried out with ferrous ions produced by electrolytically reducing ferric ions. Three end-point detection methods were employed and compared with each other: constant potential amperometry, potentiometry, and constant voltage biamperometry (a dead-stop method). The last one was found to provide high accuracy in the coulometric titration of potassium dichromate. Solution form samples were also measured to confirm the possible existence of chromium(III) in potassium dichromate by both coulometric titration and ion-exchange chromatography with inductively coupled plasma time-of-flight mass spectrometry. |
| |
Keywords: | |