首页 | 本学科首页   官方微博 | 高级检索  
     检索      


A study of retention and overloading of basic compounds with mixed-mode reversed-phase/cation-exchange columns in high performance liquid chromatography
Authors:Davies Nicola H  Euerby Melvin R  McCalley David V
Institution:Centre for Research in Biomedicine, University of the West of England, Frenchay, Bristol BS16 1QY, UK.
Abstract:The retention and overload of bases were studied on two new mixed-mode, silica based phases possessing ionic carboxylate functionalities of different acidity embedded within a hydrophobic ligand (SiELC Primesep). At low pH, good peak shapes were obtained for small solute mass, suggesting that the mere presence of a mixed-mode hydrophobic/ionic retention mechanism is not responsible for the poor peak shape that can occur on conventional reversed-phases with ionised silanols. Somewhat inferior, but still acceptable peak shape for bases was obtained on a column containing a mixture of discrete ion exchange and reversed-phase particles (Hypersil Duet). In both types of column, the ionic sites favourably increased the capacity for ionised bases, reducing considerably the deterioration of peak shape with load observed with conventional RP columns. The combined ionic and reversed-phase interaction can give strong retention of bases under certain conditions, necessitating careful choice of stationary and mobile phase.
Keywords:
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏