Evaluation of a direct extraction/liquid scintillation counting technique for the measurement of uranium in water |
| |
Authors: | J. D. Leyba H. S. Vollmar R. A. Fjeld T. A. Devol D. D. Brown J. R. Cadieux |
| |
Affiliation: | (1) Rust Remedial Services-Clemson Technical Center, 29625 Anderson, SC, USA;(2) Department of Environmental Systems Engineering, Clemson University, 29634-0919 Clemson, SC, USA;(3) Westinghouse Savannah River Company, 29808-0001 Aiken, SC, USA |
| |
Abstract: | An extraction procedure utilizing alpha liquid scintillation was evaluated for the rapid determination of uranium in aqueous environmental samples. The extraction efficiency of the system was measured under varying chemical conditions including pH. The procedure was evaluated against a traditional radiochemical technique using both laboratory prepared control samples and actual groundwater. Finally, the possibility of obtaining isotopic information from the liquid scintillation spectra was also investigated using a curve fitting routine. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|