Studies on the Electrochemical Behavior of Bleomycin at a Co/GC Ion Implantation Modified Electrode and Its Application |
| |
Authors: | HU Jing-bo LI Qi-long |
| |
Affiliation: | Department of Chemistry, Beijing Normal University, Beijing 100875, P. R. China |
| |
Abstract: | In a 0.10 mol/L Hac-NaAc buffer solution(pH= 4.59), a sensitive reduction peak of bleomycin was observed by linear sweep voltammetry at a Co/GC ion implantation modified electrode. The peak potential was -0.73 V(vs. SCE). The peak current was proportional to the concentration of bleomycin over the range of 5.0 × 10-8-1.0 × 10-6 and 1.0 × 10-6-1.0 × 10-5mol/L with a detection limit of 2.0 × 10-8 mol/L. The electrochemical behavior of the reduction peak of bleomycin at the Co/GC modified electrode was studied by linear sweep and cyclic voltammetry and applied to the determination of bleomycin in urine. This method is simple,rapid and reliable. The reduction process is quasi-reversible. The experiments of AES and XPS showed that Co was surely implanted into the surface of GCE and the depth distribution of Co was ingood agreement with Gooses normal distribution; the implanted Co at GCE improved the electrocatalytic activity. |
| |
Keywords: | Keywords Co/GC ion implantation modified electrode Electrochemical behavior Linear sweep voltammetry Bleomycin |
本文献已被 CNKI 万方数据 等数据库收录! |
| 点击此处可从《高等学校化学研究》浏览原始摘要信息 |
|
点击此处可从《高等学校化学研究》下载全文 |
|