首页 | 本学科首页   官方微博 | 高级检索  
     


Characterization of high purity gallium
Authors:Gerhard Kudermann  Karl-Heinz Blaufuß
Affiliation:(1) VAW Vereinigte Aluminium-Werke AG, Leichtmetall-Forschungsinstitut, Postfach 2468, D-5300 Bonn l, Federal Republic of Germany
Abstract:Multi-step wet analytical procedures were tested on gallium samples of different purity grades. Results obtained were compared with those of spark source mass spectrometry (SSMS) and glow discharge mass spectrometry (GDMS). It was found that multi-step procedures, as applied, with preconcentration factors of 200 are suitable to determine most of the detectable elements in high purity gallium in the ng/g-range. The results thus obtained agree well with those obtained by mass spectrometry. The sensitivity of the multi-step and mass spectrometry methods is not sufficient to detect traces of the investigated elements in super purity gallium, i.e. better than 6 N. These qualities can be differentiated, however, by single crystal resistivity measurement.
Keywords:comparative analyses  high purity gallium
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号