Abstract: | An automatic sampling device has been developed for capillary zone electrophoresis. The sample is introduced to the column electrokinetically by rapidly exchanging buffer for sample in a narrow channel drilled in a plexiglass block. The autosampler is capable, under computer control, of performing multiple sample injections from a large volume sample source (such as a reaction vessel) as the sample concentration changes, and thus presents the possibility of analyzing time-varying processes by CZE. Peak area reproducibility in electropherograms obtained after use of the sampler is less than 1% and efficiency is more than 2.2 × 105 theoretical plates. |