Calibration of a germanium well-detector using 60Co: The effects of the angular correlation of the two gamma rays emitted in cascade,quantified with Monte Carlo simulations
Institution:1. Departamento de Aceleradores, Instituto Nacional de Investigaciones Nucleares, Carr. México-Toluca S/N, Ocoyoacac, Estado de México 52750, Mexico;2. Instituto de Física, Universidad Nacional Autónoma de México, A.P. 20-364, México, D.F. 01000, Mexico;3. Centro de Ciencias de la Atmósfera, Universidad Nacional Autónoma de México, A.P. 20-364, México, D.F. 01000, Mexico;1. Department of Materials, Physics and Energy Engineering, Nagoya University, Nagoya 464-8603, Japan;2. Institute of Health Biosciences, The University of Tokushima Graduate School, Tokushima 770-8509, Japan;3. Radioisotope Research Center, Nagoya University, Nagoya 464-8602, Japan
Abstract:Radiation sources of 60Co are commonly measured by means of HPGe gamma spectrometers, either as unknown sources or as calibration sources. However, the two gamma rays that are emitted by this nuclide, at 1.17 MeV and 1.33 MeV respectively, follow each other so rapidly that in the cases where both photons interact with the crystal, the electronics will record a single additive pulse. This is a cascade or coincidence summing effect. Such effect induces in the gamma spectra a “sum-peak”, appearing at 2.5 MeV on the energy axis, generated by the pair of photons which have both been entirely absorbed by the detector. Also, the second photon is correlated for direction to the first one, i.e., it will not be emitted at random, with an isotropic probability. Then the question arises of to what extent the two effects, cascade and correlation, might affect the count rates for the three peaks. Various answers have already been published, but without practical solutions. In this context, the present work was devoted to further explore the question by means of Monte Carlo simulation, in the case of a welltype detector. As a result, inside the well the sum peak only is not affected and it allows accurate evaluation of the detector efficiency at 1.25 MeV. Outside the well, near the detector, none of the three peaks can be directly used for efficiency evaluation, unless the relevant corrections can be evaluated. At a distance from the detector, the two single peaks can be used, but with the drawback of a low efficiency.
Keywords:Germanium-detector  Cascade effect  Angular correlation  Monte Carlo simulation
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号