Characterization of miniature RAD-HARD silicon diodes as dosimeters for small fields of photon beams used in radiotherapy
Institution:1. Instituto de Pesquisas Energéticas e Nucleares IPEN-CNEN/SP, Av. Professor Lineu Prestes, 2242, 05508-000 São Paulo, SP, Brazil;2. Hospital Sírio Libanês, Rua Dona Adma Jafet, 91, 01308-050 São Paulo, SP, Brazil;3. Pontifícia Universidade Católica de São Paulo PUC/SP, Rua Marquês de Paranaguá, 111, 01303-050 São Paulo, SP, Brazil;4. Helsinki Institute of Physics and Laboratory of Physics, Helsinki University of Technology, P. O. Box 1100, FIN-02015 HUT, Espoo, Finland;1. National Physical Laboratory, Hampton Road, Teddington, Middlesex TW11 0LW, UK;2. Urenco Ltd, Almelo, The Netherlands;1. Laboratoire Commun de Métrologie (LNE-CNAM), Laboratoire National de Métrologie et d’Essais (LNE), 1 Rue Gaston Boissier, 75015 Paris, France;2. Ecole Polytechnique Fédérale de Lausanne, Rue de la Maladière 71B, CH-2002 Neuchâtel, Switzerland;3. Université Paris-Est, laboratoire ESYCOM, ESIEE-Paris, Cité Descartes, 2 Bd Blaise Pascal, 93162 Noisy-Le-Grand Cedex, France
Abstract:The dosimetric response characterization and beam data acquisition performance of a miniature Float Zone (FZ) silicon diode for photon beams was investigated using Novalis TX linear accelerator (Varian Medical Systems®). In all measurements the unbiased diode operated in a short-circuit mode, connected to the input of a Keithley 6517B electrometer using a water phantom. For photon beams of the 6 and 15 MV the results presented good repeatability (coefficient of variation ≤1.6%), measured through switching on/off the photon beams. Moreover, the diode showed a quite linear response, given by the charge versus absorbed dose, with charge sensitivities higher than 6.9 nC/Gy. The output factor, percentage depth dose profile (PDD) and transversal dose profile (TDP) were also measured in a water phantom. For small field sizes, the output factor values using the FZ diode were compared with measurements obtained with a SFD (Stereotactic Field Diode) commercial diode and the differences were 5.4%, 2.5% and 1.3% for the field sizes of 1 × 1, 2 × 2 and 3 × 3 cm². For larger field sizes (≥4 × 4 cm²), the maximum difference found was 0.7% in comparison with values obtained with a CC13 ionization chamber. Thus, the result demonstrates that the unshielded FZ diode has the potential to be used for measuring of, as it performed acceptably well for both small and large field sizes. The TDP experimental results obtained with the FZ diode for field sizes of: 1 × 1 cm2, 2 × 2 cm2 and 4 × 4 cm2 are in agreement with experimental results acquired with several commercial detectors. Through the TDP study, the comparison of the field penumbra size confirmed the excellent spatial resolution of the miniature diode. However, the PDD study, requires further investigation.
Keywords:FZ diode  Radiotherapy photon beam  Small fields  Dosimetry
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号