Micro-electrospray with stainless steel emitters |
| |
Authors: | Shui Wenqing Yu Yanling Xu Xuejiao Huang Zhenyu Xu Guobing Yang Pengyuan |
| |
Affiliation: | Department of Chemistry, Fudan University, Shanghai 200433, China. |
| |
Abstract: | The physical processes underlying micro-electrospray (micro-ES) performance were investigated using a stainless steel (SS) emitter with a blunt tip. Sheathless micro-ES could be generated at a blunt SS tip without any tapering or sanding if ESI conditions were optimized. The Taylor cone was found to shrink around the inner diameter of the SS tubing, which permitted a low flow rate of 150 nL/min for sheathless microspray on the blunt tip (100 microm i.d. x 400 microm o.d.). It is believed that the wettability and/or hydrophobicity of SS tips are responsible for their micro-ES performance. The outlet orifice was further nipped to reduce the size of the spray cone and limit the flow rate to 50-150 nL/min, resulting in peptide detection down to attomole quantities consumed per spectrum. The SS emitter was also integrated into a polymethylmethacrylate microchip and demonstrated satisfactory performance in the analysis and identification of a myoglobin digest. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|