CMOS arrays as chemiluminescence detectors on microfluidic devices |
| |
Authors: | Eunice R G O Rodrigues Rui A S Lapa |
| |
Institution: | 1. REQUIMTE, Servi?o de Química-Física, Faculdade de Farmácia da Universidade do Porto, Rua Aníbal Cunha 164, 4099-030, Porto, Portugal
|
| |
Abstract: | A simple, low-cost process to integrate complementary metal oxide semiconductor array detectors (CMOSAD) for chemiluminescence
is presented, evaluated, and applied to the determination of nitrite in ground water samples. CMOS arrays of different brands
(obtained from commercial image sensors) were adapted as chemiluminescence detectors on microfluidic devices. The performance
of the CMOSADs was evaluated in the visible zone of the spectrum using a tungsten halogen lamp as light source. Intrinsic
parameters assessed included signal stability, spectral response, dark current, and signal-to-noise ratio. Thereafter, the
CMOSADs were integrated on microfluidic devices and their performances in quantitative analysis were assessed with the chemiluminometric
reaction of hydrogen peroxide with luminol, catalyzed with hexacyanoferrate (III). The parameters assessed were sensitivity,
linear range, detection limit, reproducibility, correlation coefficient of the calibration curves, and baseline drift during
measurements. The CMOSAD with the best performance was selected to assess the applicability of the developed microfluidic
devices with the integrated detector. The microfluidic system permitted the determination of nitrite with both good precision
and good recovery values in the analysis of ground water samples. Integration was easily achieved and enabled the development
of a simple, low-cost, and feasible alternative to conventional detectors. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|