首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Atomic absorption determination of Cd, Co, Cu, Fe, Ni in viscous aqueous solutions of poly(ethyleneimine) using a high-pressure flow and nebulization system for sample introduction
Authors:H Berndt  G Schaldach  B Y Spivakov and V M Shkinev
Institution:(1) Institut für Spektrochemie und angewandte Spektroskopie, P. O. Box 10 13 52, D-4600 Dortmund 1, Federal Republic of Germany;(2) Vernadsky Institute of Geochemistry and Analytical Chemistry, Academy of Sciences, V-334 Moscow, Russia
Abstract:Summary The influence of increasing concentration of water-soluble polymers on the analytical signals for Cd, Co, Cu, Fe, Ni in flame atomic absorption spectrometry with hydraulic high-pressure nebulization (HHPN) and pneumatic nebulization (self uptake mode) has been studied using aqueous solutions of poly(ethyleneimine) (PEI) as examples. For a solution of 50 g/l polymer, HHPN sample introduction leads to a 5.0 (Cd), 8.8 (Co), 17 (Cu), 7.1 (Fe) and 22 (Ni) fold higher sensitivity compared to pneumatic nebulization. Even at polymer concentrations of more than 100 g/l HHPN operation is still possible without any problems, whilst such viscous solutions cannot be transferred into an aerosol by a pneumatic nebulizer.
Keywords:
本文献已被 SpringerLink 等数据库收录!